HG-1 Argon Wavelength Calibration Source Ocean Optics

The HG-1 is a spectral wavelength calibration source specifically designed for UV-VIS spectrometers like our Jaz, USB2000, and HR series. The Hg-1 produces low-pressure mercury and argon emission lines from 250-1000 nm for use in performing fast, reliable spectrometer wavelength calibrations. The spectral emission lines are printed on the lamp's housing. Use a 50um UV-VIS fiber with SMA connector for best results

 

Engineering Specifications HG-1
Source: Low-pressure gas discharge lines of mercury and argon
Wavelength Range: 273 – 1704 nm (253 – 922 nm first order emission lines)
Dimensions: 12.5 x 7.0 x 2.6 cm
Weight: 40 g
Power consumption: 250 mA at 12 VDC
Power requirements: 12 VDC wall transformer (included) or
9 VDC battery (additional)
Internal voltage: 600 V at 30 kHz
Bulb life: ~ 3,500 hours (at 20 mA)
Amplitude stabilization: ~1 minute
Aperture: 3 mm
Connector: SMA 905

 

Strong Mercury Emission Lines (nm)

 

253.652 404.656
296.728 407.783*
302.150 435.833
313.155 546.074**
334.148 576.960
365.015 579.066

* This spectral line will not be evident with spectrometers configured with 300- or 600-lines/mm gratings.

** With spectrometers that have 1200-, 1800-, 2400- or 3600-lines/mm gratings, spectral lines will be evident at 576.96 nm and 579.07 nm.

 

Strong Argon Emission Lines (nm)

 

696.543 800.616*
706.722 811.531
714.704 826.452
727.294 842.465
738.398 852.144
750.387 866.794
763.511 912.297
772.376 922.450
794.818
 

Strong Argon Emission Lines (nm)

 

696.543 800.616*
706.722 811.531
714.704 826.452
727.294 842.465
738.398 852.144
750.387 866.794
763.511 912.297
772.376 922.450
Mercury Argon Hg-A discharge tube wavelength calibration standard
Click To Enlarge
  • Item #: HG-1
  * Marked fields are required.
Qty*
Price $553.00

Related Items

Reviews (0) Write a Review
No Reviews. Write a Review