F Style O2 Sensor Dip Coated

Fiber Optic sensor consists of 1000um fiber, polished at 45 degrees and coated with one of 3 chemistries. AP is for general purpose and exhibits zero drift.  HCR is for use in hydrocarbons, and MR is for use in applications where faster rsponse is required (call us if you want to use MR)

The optical fiber is terminated in an SMA connector, and a SEOX-BB1000-X bifurcated fiber and splice bushing is required.

Probes are factor calibrated 0 - 20.9% O2 in gas and 10 - 40C temperature range. Broader ranges or calibrations in specific solvents are available at an additional charge.

The NEOFOX-TP thermistor assembly is reccomended for non-temperature regulated applications.

These probes are designed to work with NEOFOX or NEOFOX-SPORT electronics but should work fine in other phase fluorometer type systems.

 

 

AP formulation

Polystyrene matrix

HCR formulation

Sol Gel matrix

MR formulation

Sol Gel matrix

Response Rate

T90 = 3 sec

T90 = 1 sec

T90 = 1 sec

Dynamic Range (gas)

0 – 760 mmHg

0 – 100% (at 1 atm)

0 – 760 mmHg

0 – 100% (at 1 atm)

0 – 760 mmHg

0 – 100% (at 1 atm)

Dynamic range (DO)

0 – 40 ppm in H2O

0 – 40 ppm in H2O

0 – 40 ppm in H2O

Resolution (gas)

0.04 mmHg

0.005% (at 1 atm)

0.4 mmHg

0.05% (at 1 atm)

0.4 mmHg

0.05% (at 1 atm)

Resolution (DO)

0.002 ppm in H2O

0.002 ppm in H2O

0.002 ppm in H2O

Accuracy

5% of reading

5% of reading

5% of reading

LDL (gas)

0.04 mmHg

0.005% (at 1 atm)

0.4 mmHg

0.05% (at 1 atm)

0.4 mmHg

0.05% (at 1 atm)

LDL (DO)

0.002 ppm in H2O

0.002 ppm in H2O

0.002 ppm in H2O

Drift rate

No measureable long term drift, continuous operation

0.0001 % /hr in gas

0.0001 % /hr in gas

Sensor Operating Temperature

-50 to 100C

-50 to 200C

-50 to 200C

Useful Lifetime

3 years

3 years

3 years

Compatibility

General purpose gases and aqueous solutions

Organic solvents

General purpose gases and aqueous liquids

Water Vapor Interference

no

yes

no

Formats

Dip coated fibers

Spin coated patches & probes

Dip coated fibers

Spin coated patches & probes

Dip coated fibers

Spin coated patches & probes

Comments

Best general purpose probe, longer fluorescence decay yields more resolution, zero drift means stable calibrations

Used for organic solvents and gases with organic vapors.  This sensor does respond to water vapor as well as O2

General purpose but with faster response rate than AP chemistry. Does exhibit long term photo-bleaching drift and has less resolution.

 

O2 oxygen sensor 1000um dip coated optical fiber
Click To Enlarge
  • Item #: SEOX-F
  * Marked fields are required.
Qty*
Price $400.00

Related Items

Reviews (0) Write a Review